074-578-P1A 44
TVOC concentration over
instrument operating range
Chamber temperature too high
Hydrogen fuel cylinder has 10
hours of use remaining
Hydrogen fuel cylinder has 5 hours
of use remaining
Sending log data to
the PC
Instrument sending log data to PC
Zero value too high during
calibration
Span gas too low during
calibration
Table 5. System Alerts
System Alerts and Alarms
While operating the instrument, system alerts can occur. To accurately identify the source of the
alarm, each type of alarm has been given a unique status.
In addition to the status, the DataFID also has an audio alarm and an alarm LED. A soft key is
used for acknowledging alarms and is named “Accept”. To clear the alarm, press the “Accept”
key. Once acknowledged, the alarm indicators are cleared. The alarm status will remain until the
alarm condition clears.
The DataFID updates the instantaneous concentration once every second. Following every
update, the instantaneous concentration is compared to the selected compounds high alarm
level, and if exceeded, an alarm is triggered.
During calibration, all alarms are disabled. Once the calibration is complete, the alarms are re-
enabled.
User Interface – Basic Menu
The DataFID is designed for ease of use with a logically organized internal menu structure/user
interface. The DataFID’s Main Menu options are shown in Figures 13, 14, 15, and 16.
The DataFID has three soft keys under the graphic LCD display which always show available
functions of the soft keys in any screen.