LD Protocol 37
321 141
DIP switch
MSB
RUINT8
DIP switch setting of the MSB:
Bit7: S171, switch 4
Bit6: S171, switch 3
Bit5: S171, switch 2
Bit4: S171, switch 1
Bit3: S170, switch 4
Bit2...0: not used,always 0
322 142
Field bus status
word
RUINT16
Status word for Bus modul
refer to Bus module documentation
323 143
SW version bus
modul
R UINT8[3] SW version bus modul
324 144
Bus module
fieldbus type
RUINT16
Bus module fieldbus type.
Refer to AnybusCC specification for
enumeration.
325 145
Serial number
plug-in unit bus
modul
R UINT8[4] Serial number plug-in unit bus modul
326 146
Field bus
address
RUINT8
Fiedbus address
Refer to AnybusCC specification for
enumeration.
327 147
Field bus baud
rate
RUINT8
Baud rate at field bus
Refer to AnybusCC specification for
enumeration.
328 148
Exception code
bus modul
R UINT8 Exception code bus module
329 149
Error counters
bus module
R UINT16[4]
Error counters bus module
Index:
0: Discarded commands
1: Discarded Responses
2: Serial Reception errors
3: Fragmentation errors
330 14A
Bus module
state
RUINT8
State of bus module
see Enumarations
331 14B
Field bus
address
nominal value
R/W UNIT8
Fieldbus address nominal value. Refer to
AnybusCC specification for enumeration.
385 181
Trigger [mbar*l/
s]
R/W FLOAT[4] 1E-12, 1E-5, 1E3 Trigger in mbar*l/s
390 186
Test leak
extern vacuum
[mbar*l/s]
R/W FLOAT
1E-9, 9.9E-2,
9.9E-2
Test leak extern for vacuum mode in
mbar*l/s
392 188
Test leak
extern sniff
[mbar*l/s]
R/W FLOAT
1E-7, 9.9E-2,
9.9E-2
Test leak extern for sniff mode in
mbar*l/s
394 18A
Testleak intern
[mbar*l/s]
R/W FLOAT
1E-7, 9.9E-2,
9.9E-2
Testleak intern in mbar*l/s
401 191
Operation
mode
R/W UINT8 0, 0, 1
0 = VACUUM
1 = SNIFF
Comm
and dez
Com
mand
hex
Name R/W Data type
Min-, Def.-, Max-
value
Meaning