8 ASCII Protocol
Each command is answered with the requested data, „ok“ or „EXX“ (in case of an
error). For a list of all error messages 2.6. The transmission can be cancelled and the
receive-buffer will be cleared with ESC (ASCII code 27dec/1Bhex), ^C (ASCII code
3dec/03hex) or ^X (ASCII code 24dec/18hex).
Some commands can be used as queries, some can be used to set menu parameter
and some can be used for both. A query is marked by a „?“ (ASCII code 63dec/
3Fhex) after the command; for setting data the command has to be followed by the
new value to be set.
Parameter can be Boolean or numerical:
Notice Always use a point as the decimal marker. If a comma is used during
numerical data entry, the conversion of the number is cancelled at this
point and only the integer part of the number will be used.
Timing recommendations for the PC/PLC - Program:
Sample rate > 100 ms
Timeout between request to and answer from LDS3000: 1500 ms
After sending a command the answer must be waited for before sending a new
command. Otherwise the receive buffer may be overwritten.
2.4 Commands
<b> Boolean 0 / 1 or OFF / ON
<No> Numeric representation format: integer, real (15.6) or exponential (4.5
-7
)
Format: [space] [sign] [ddd] [.] [e[sign]ddd] (d:digit)
Relates to
LD cmd. no.
Read /
Set
*CLS Clear Error 5 S
*IDN Identification
:CRC check sum 320 R
:DEVice name of instrument 301 R
:VERsion software version MSB 310 R
:SERial serial-number leak detector R
:TURBO software version TMP controller 315 R
:DIP1 MSB DipSwitch 1 (binär) 321 R
:DIP2 MSB DipSwitch 2 (binär) 321 R
:CUversion software version control unit 314 R
:IOversion software version I/O modul 313 R
:TCHARDware hardware version TMP controller 316 R
:TCNAME TMP controller name 317 R
:BMVersion software version Bus modul R
:BMSerial serial-number Bus modul R
:BMNETType Bus module network type R
*STATus
status of LDS2010 (
ACCL, STBY, MEAS, CAL, ERROR,
EMIOFF)
Status
word
R
Vista de pagina 7
1 2 3 4 5 6 7 8 9 10 11 12 13 ... 55 56

Comentarios a estos manuales

Sin comentarios

Audio-Technica Mixer audio manuali utente

Partendo da A

Cliccare su un alfabeto di seguito per far scorrere la lista completa dei modellif models starting with that letter

Modelli Tipo di Documento
AT-MX351a SmartMixer
Manuale Utente    product catalogue I europe I €, 67 pagine