60 Operation
jina85e 5.fm Technical Handbook(1112)
Interfaces RS232 Protocol Binary
The binary protocol can be used to communicate with T-Guard™ Leak Detection
Sensor by software, written by a programmer. This method of communication is very
fast. It is, for example, the fastest way to read a leak rate from the T-Guard™ Leak
Detection Sensor.
Interfaces RS232 Protocol Printer Auto
After each measurement, the T-Guard™ Leak Detection Sensor sends data via
RS232.
Date, Time, Parameter set name, Measurement mode, Leak rate, Trigger 1, [Trigger
2].
Interfaces Select PLC inputs
In this screen, you can select which functions can be called from the 6 inputs.
For default settings refer to section 4.6.2.
Interfaces Select PLC-outputs
In this screen you can choose which information will be present on the 8 outputs.
For default settings refer to section 4.6.4.
Interfaces Scaling Analog Output Analog Output 1 .. 2
In this screen, you can set the upper limit and the volts/decade for the output on Pin
1 and 2.
Interfaces Scaling Analog Output Analog Output 3 .. 4
In this screen, you can set the upper limit and the volts/decade for the output on Pin
3 and 4.
InterfacesAnalog Output
In this screen you can choose which information will be present on the two analog
outputs.
Possible values: Off, Pressure, LR Mant, LR Exp, LR Lin, LR Log., Current Lin,
Current Log.
Default setting: 1..2: LR Exp, 3..4: LR Log.
Interfaces Scaling Analog Output
In this screen, you can select how the voltage signal of the analog output has to be
interpreted, when "LR Log." or "Current Log." have been selected.