
PM-700 SERIES PLATING MONITOR
49
4.8.4.3 Remote Control
4.8.4.3.1 Upload Parameter Store (Instruction 10h)
This operation outputs the entire parameter store from the monitor to the host
computer. All the parameters are output with the most significant byte first. The
parameters are output in the following order:
Table 7 Upload Parameter Store
No. Position Description Length
(Bytes)
Decimal
Point Pos.
1 1 Plate endpoint 3 *
2 4 Preplate endpoint 3 *
3 7 Minimum rate alarm 3 *
4 10 Maximum rate alarm 3 *
5 13 Preplate rate 3 *
6 16 Dwell time in % 3 3
7 19 Interval time (seconds) 3 N/A
8 22 Deplate rate 3 *
9 25 Tooling factor % 3 1
10 28 Material density (gm/cm
3
) 3 1
11 31 Acoustic imp. [10
5
gm/(cm
2
sec)] 3 1
12 34 Electrochemical eqv. (gm/Amphr) 3 1
13 37 DAC thickness range 3 *
14 40 DAC low rate 3 *
15 43 DAC high rate 3 *
16 46
Simulate rate (µg/cm
2
/min)
3 1
17 49 Plate current density 3 *
18 52 Deplate thickness % 3 3
19 55 Electrode area in cm
2
3 1
20 58 Plating material 1 N/A
21 59 Sample (On|Off) 1 N/A
22 60 Bath type(Electro|Elecless) 1 N/A
23 61 Main display 1 N/A
Comentarios a estos manuales