1 - 16
PN 074-156N
Front Load Single and Dual Sensors Operating Manual
Water Quality . . . . . . . . . . . . . . . . . . Coolant should not contain chlorides as
stress corrosion cracking may occur.
Extremely dirty water may result in loss of
cooling capacity.
Do not allow water tubes to freeze. This may happen if the
tubes pass through a cryogenic shroud and the flow of
fluid is interrupted.
Air Pressure . . . . . . . . . . . . . . . . . . . 70 psi (gauge) {85 psi (absolute)}
(5.8 bar (absolute)) [584 kPa (absolute)]
(minimum)
80 psi (gauge) {95 psi (absolute)}
(6.5 bar (absolute)) [653 kPa (absolute)]
(maximum)
Do not exceed 100 psi (gauge) {115 psi (absolute)}
(7.9 bar (absolute)] [791 kPa (absolute)].
Connection to excessive pressure may result in personal
injury or equipment damage.
1.6.4 Dual Sensor Drawings
The following Dual Sensor Outline and Assembly Drawings provide dimensions
and other relevant data necessary for planning equipment configurations.
Figure 1-11. . . . . Dual Crystal Sensor Outline
Figure 1-12. . . . . Dual Crystal Sensor Assembly (PN 750-212-GX)