PN 074-156N
Front Load Single and Dual Sensors Operating Manual
1.8.2 Specifications for CF40 (2-3/4 in. ConFlat)
Mounting . . . . . . . . . . . . . . . . . . . . . Mates with 7 cm (2-3/4 in.) ConFlat® type
flanges with 3.49 cm (1.375 in.) ID
(minimum)
Ultra-Torr O-ring compression fitting terminations
Maximum operational
environment temperature
with no water . . . . . . . . . . . . . . . . . . 165°C
Maximum operational
environment temperature
with minimum water flow . . . . . . . . . 300°C
Materials. . . . . . . . . . . . . . . . . . . . . . 304 stainless steel, Teflon, glass, brass,
beryllium copper, VITON®
Welded terminations
Maximum operational
environment temperature
with no water . . . . . . . . . . . . . . . . . . 165°C
Maximum operational
environment temperature
with minimum water flow . . . . . . . . . 450°C
Materials. . . . . . . . . . . . . . . . . . . . . . 304 stainless steel, Teflon, glass, brass,
beryllium copper